YZ200 Coating Thickness Gauge

Measurement Range   0—1250um or 0—50mil


 Accuracy± [(1%~3%)H+1um]  NoteH is thickness reading

 powers down after 60 seconds of no activity.

KS200 Coating Thickness Gauge

Operation Guide:

1、Power-on and Warm-up
Keystroke once;Wait until  READY! is displayed on the LCD. Now the gauge is ready for measurements.
2、Zeroing
Keystroke twice; when  BASE ZERO  is displayed, place flat on Fe or Al base; hold steady. After measurement is obtained,  DONE  is shown and gauge beeps. Then lift gauge.
Note: Please conduct zeroing process after changing base. Zeroing process could be repeated for better results.
3、Unit Change
Keystroke three times; convert from ums to mils or vice versa.
4、Reset
Keystroke four times,Wait until  RESET  is displayed; after gauge resets, users can conduct zeroing process if necessary.

Materials

Gauge    Reference Fe and Al Base
Certified Plastic Shim   Battery etc.YZ200 Coating Thickness Gauge