YZ100 Coating Thickness Gauge

 

1.Measurement Range:   0—1250um or 0—50mil

  1. Accuracy:± [(1%~3%)H+1um]  Note:H is thickness reading
  2. powers down after 60 seconds of no activity.

    1、Power-on
    Keystroke once;Wait until  READY! is displayed on the LCD. Now the gauge is ready for measurements.
    2、Zeroing
    Keystroke twice; when  BASE ZERO  is displayed, place flat on surface; hold steady. After measurement is obtained,  DONE  is shown and gauge beeps. Then lift gauge.
    Note: Please conduct zeroing process after changing base. Zeroing process could be repeated for better results.
    3、Two-Point User Calibration
    Keystroke three times, enter into the process of tow-point user calibration; when
    USER CAL BASE  is shown on the LCD, please conduct a measurement on the base first; when
    S1  is shown on the LCD, place Shim S1 on the base, conduct a measurement on shim S1 and lift gauge; when
    S2  is shown on the LCD, place Shim S2 on the base, conduct a measurement on shim S2 and lift gauge; When  DONE  is displayed, it is ready for use.
    Note: The two-point user calibration process could be repeated for better results.
    4、Unit Change
    Keystroke four times, convert from ums to mils or vice versa.
    5、Reset
    Keystroke five times, gauge resets.